How do titanium and Ti6Al4V corrode in fluoridated medium as found in the oral cavity? An in vitro study - CentraleSupélec Access content directly
Journal Articles Materials Science and Engineering: C Year : 2015

How do titanium and Ti6Al4V corrode in fluoridated medium as found in the oral cavity? An in vitro study

Abstract

The purpose of this work was to evaluate the corrosion of commercially pure (CP) titanium and Ti6Al4V in vitro at different F− concentrations regularly found in the oral cavity by using different electrochemical tests and surface analysis techniques. electrochemical impedance spectroscopy (EIS), open circuit potential (OCP) and potentiodynamic polarization tests were associated to advanced characterization techniques such as SEM, EDS, AFM, ICP-MS and XPS. OCP tests revealed a higher reactivity of both CP titanium and Ti6Al4V at 12,300 ppm F− concentration than that recorded at 227 ppm F−. Also, a significant decrease of the corrosion resistance of both materials was noticed by EIS in fluoride solutions. Material loss caused by corrosion was noticed on titanium surfaces by SEM and AFM in the presence of high F− concentration. CP titanium degraded by pitting corrosion while Ti6Al4V suffered from general corrosion showing micro-cracks on surface. Furthermore, a high release of metallic ions from the test samples after immersion at high F− concentrations was detected by ICP-MS, that can be potentially toxic to oral tissues.

Dates and versions

hal-01234208 , version 1 (26-11-2015)

Identifiers

Cite

Júlio C.M. Souza, Sandra L. Barbosa, Edith A. Ariza, Mariana Henriques, Wim Teughels, et al.. How do titanium and Ti6Al4V corrode in fluoridated medium as found in the oral cavity? An in vitro study. Materials Science and Engineering: C, 2015, 47, pp.384-393. ⟨10.1016/j.msec.2014.11.055⟩. ⟨hal-01234208⟩
80 View
0 Download

Altmetric

Share

Gmail Facebook Twitter LinkedIn More